A late preterm infant who is at risk of hypoglycaemia should be screened with regular monitoring of the capillary glucose concentrations.
Near patient testing devices tend to be less accurate in the lower range, especially < 2.0mmol/l [1] and therefore all low values (≤2.6mmol/L) require confirmation using blood gas analysis as this is considered the gold standard for measuring blood glucose.
Hand held glucometers should meet ISO standards (ISO15197:2013). If a handheld glucometer is used, low levels must be confirmed using an accurate method as cot-side monitors may be inaccurate in the lower ranges and require checking using a True Blood Glucose (TBG) to guide therapy. A TBG can be obtained by sending a formal laboratory sample but significant delays can occur in obtaining a result, alternatively a TBG can also be obtained from a blood gas analyser, where available, as these are equally reliable [2]. All units must ensure they have readily accessible methods for accurate measurement of a TBG . Each unit must be aware of the characteristics of any near patient testing device used in their hospital.
Local Arrangements for Confirming Blood Glucose <2.6mmol/L
GG&C - GG&C Maternity and Neonatal units use the Precision Exceed Pro meter©. For this device glucose values 2.6 - 3.0mmol/l may prove to be <2.6mmol/l when a TBG is obtained from a blood gas machine or laboratory testing. Therefore a TBG should be obtained for:
- Symptomatic babies with values below 3.0mmol/l (see section on symptoms of hypoglycaemia)
- Asymptomatic babies with 2 values < 2.6mmol/l OR any value<2.0mmol/l