A baby at significant risk of hypoglycaemia, or who has symptoms which may be secondary to hypoglycaemia, should be screened with regular monitoring of the capillary glucose concentrations.
Near patient testing devices tend to be less accurate in the lower range, especially < 2.0mmol/l7and therefore all low values (≤2.6mmol/L) require confirmation using blood gas analysis as this is considered the gold standard for measuring blood glucose.
Hand held glucometers should meet ISO standards (ISO15197:2013) and have CE marking as described in the BAPM Framework. If a handheld glucometer is used, low levels must be confirmed using an accurate method as cot-side monitors may be inaccurate in the lower ranges and require checking using a True Blood Glucose (TBG) to guide therapy A TBG can be obtained by sending a formal laboratory sample but significant delays can occur in obtaining a result, alternatively a TBG can also be obtained from a blood gas analyser, where available, as these are equally reliable8. All units must ensure they have readily accessible methods for accurate measurement of a TBG . Each unit must be aware of the characteristics of any near patient testing device used in their hospital.
Local Arrangements for Confirming Blood Glucose <2.6mmol/L
GG&C - GG&C Maternity and Neonatal units use the Precision Exceed Pro meter©. For this device glucose values 2.6 - 3.0mmol/l may prove to be <2.6mmol/l when a TBG is obtained from a blood gas machine or laboratory testing. Therefore a TBG should be obtained for:
- Symptomatic babies with values below 3.0mmol/l (see section on symptoms of hypoglycaemia)
- Asymptomatic babies with 2 values < 2.6mmol/l OR any value<2.0mmol/l