|
There are 2 potential set-ups for the transport ventilator and each requires a slightly different arrangement for checking of the flow sensor & adaptor - these are weight banded:
<15kg: For this document we will term this group NeoPaeds, |
|
This set-up includes: |
Pink expiratory flow valve (comes separately from other items below) Blue/White vent tubing Blue/White flow probe Dumbbell shaped adaptor |
≥15Kg: This group uses standard set up for adults |
|
This set-up all comes in the same bag & includes: |
Blue/grey expiratory flow valve White dual lumen vent tubing Blue/White flow probe Funnel shaped adaptor |
This table summarises the different set ups:
|
Valve (type/colour) |
Flowsensor |
Circuit |
Adaptor for pre-use check |
NeoPaeds |
Neonatal (pink) |
Neonatal/paeds |
Paediatric |
Dumbbell |
Adult |
Adult (Blue or grey - white/plastic) |
Adult |
Adult |
Funnel |
In-line suction should be used at all times where possible (ETT ≥ 4.0)
Please also read section 9.1 if further is required for highly infectious disease transfers
All necessary equipment is stored by the Hamilton-T1 in the PICU Store Room
In order to maintain staff safety we should adjust the circuit set-up with the following filters
Attach filter to expiratory port (Intersurgical Clear Guard Midi - ref 1644000)
<4.0 ETT patient-end of circuit should be set-up as detailed below
ETT ≥4.0 should be set up as follows, with appropriate sized ventilator tubing & flow sensor & in-line suction as per the patients weight
A - Attach connector to inspiratory port (Intersurgical Connector 22M-22F- ref 1961000S)
B - Attach filter to expiratory port (Intersurgical Clear Guard Midi - ref 1644000)
In-line suction should be used at all times where possible (ETT ≥ 4.0)
ETCO2 should be placed “above” the pink Pharma HME-F filter (this allows VT up to 900ml) as detailed below
As routine we should also ensure the T-piece has the bacterial/viral filters (Intersurgical HME-F 14410000) attached as shown and gas supply to the T-piece should be switched off when not in use
These filters should be changed every 24 hours
Use small ETCO2 if ETT is 4.5 or less and large ETCO2 if ETT 5.0 or above
NeoPaeds Flow Sensor Check |
Adult Flow Sensor Check |
NB Remember to remove the “pre-use check” adaptor prior to use on a patient
O2 test check if needed – we don’t do this routinely
NB
The Pinsp is actually a DP (ie Pressure above PEEP)
Set Ti for ventilation last (as will reset if you adjust weight/height settings)
Auto-set alarms can be made only adult mode
NoxBox can be attached to the Hamilton T1 using the neo/paeds tubing only:
7.1 Access NoxBox pack | 7.2 Open contents | 7.3 Select components |
7.4 Attach funnel to iNO. Attach sampling line to delivery device | 7.5 Attach to inspiratory port (blue) of Hamilton T1 | 7.6 Remove adaptor from nCPAP pressure pack (in vent tubing pack) |
7.7 Attach adaptor between inspiratory limb (blue) and swivel vent connector | 7.8 Attach NoxBox sampling line to adaptor | |
Secure for patient transfer as noted below:
Attach onto monitor stand as noted and secure using leather strap
Connect ventilator tubing to oxygen splitter to enable blended FiO2/Air to be delivered if needed as well
Parameter |
Definition |
Apnea Backup | A function that provides ventilation after the adjustable apnea time passes without breath attempts. If "Automatic" is enabled, control parameters are calculated based on the patient's IBW. |
ETS | Expiratory trigger sensitivity. The percentage of peak inspiratory flow at which the ventilator cycles from inspiration to exhalation. |
Flow trigger | The patient's inspiratory flow that triggers the ventilator to deliver a breath. |
Gender | Sex of patient. Used to compute ideal body weight (IBW) for adults and pediatrics. |
I:E | Ratio of inspiratory time to expiratory time. Applies to mandatory breaths. |
%MinVol | Percentage of minute volume to be delivered in ASV mode. The ventilator uses the %MinVol, Pat. height, and Gender settings to calculate the target minute ventilation. |
Oxygen | Oxygen concentration to be delivered. |
Pasvlimit | The maximum pressure to apply in ASV mode. Changing Pasvlimit or the Pressure alarm limit automatically changes the other: The Pressure alarm limit is always 10 cmH2O greater than Pasvlimit. |
Pat. height | Patient height. It determines the ideal body weight (IBW), which is used in calculations for ASV and startup settings for adults and pediatric patients. |
Pcontrol | The pressure aditional to PEEP/CPAP. |
PEEP/CPAP | Positive end expiratory pressure. |
P high | The high pressure setting in APRV and DuoPAP modes. Absolute pressure, including PEEP. |
Pinsp | Pressure (additional to PEEP/CPAP) to apply during the inspiratory phase. Applies in PSIMV+ IntelliSync and NIV-ST. |
P low | The low pressure setting in APRV. |
P-ramp | Pressure ramp. Time required for inspiratory pressure to rise to the set (target) pressure. |
Psupport | Pressure support for spontaneous breaths in SPONT, NIV, and SIMV+ modes. |
Rate | Respiratory frequency or number of breaths per minute. |
Sigh | Breaths delivered at a regular interval (every 50 breaths) at a pressure up to 10 cmH2O higher than non-sigh breaths, as allowed by the Pressure alarm limit. |
Thigh | Length of time at the higher pressure level, P high, in DuoPAP and APRV modes. |
TI | Inspiratory time, the time to deliver the required gas (time to reach the operator-set Vt or Pcontrol value). |
TI max | Maximum inspiratory time for flow-cycled breaths in NIV, NIV-ST, and SPONT in neonatal modes. |
TI low | Length of time at the lower pressure level, P low, in APRV mode. |
Vt | Tidal volume delivered during inspiration in (S)CMV+ and SIMV+ modes. |
Vt/kg | Tidal volume per weight. |